User Manual
–Error message
NIMXEN020I
Owandy Radiology SAS
96
Sensor Ready
Code
Error description
370
Sensor ready lost during exposure
371
Sensor not ready
Temperature sensor
Code
Error description
500 ÷ 503
Temperature sensor reading error
Generator Board
Code
Error description
750
Generator board initialization error
751
Alarm "overvoltage kV"
752
Alarm "overload on filament" on Generator board
753
Alarm "overload anodic current"
754
Alarm "filament not OK"
755
Alarm "backup timer"
756
Alarm "PFC not OK"
757
Alarm "Brown OUT"
758
Alarm "NO X-ray"
759
Alarm "unexpected emission"
760
Alarm "NO RX button command"
761
Alarm "NO X-ray emission"
762
Bad unit status: emission flag detected unexpectedly
763
kV analog feedback out of range
764
mA analog feedback out of range
765
Filament analog feedback out of range
766
Generator board reset due to a brown out
767
Generator board reset due to low voltage detection
768
Generator board reset due to a watchdog timeout
769
Generator board reset due to a stack overflow
770
Mismatch between generator board (A2) and MCU board (A1) types (2D
/ 3D)
Keyboard
Code
Error description
850
One or more keycodes are pressed
852
Button >O< pressed during movements
DSPU
Code
Error description
1003
DSPU hardware error detected
1004
MCU
– DSPU communication error
1005
CAN Bus hardware error
1007
Flat panel temperature too high