User Manual
– Technical characteristics
NIMXEN020I
Owandy Radiology SAS
35
6.2.2
PC suggested characteristics
•
Mother board with at least one free PCI express X16 slot (mandatory for the Dual Port
Network Card provided with the machine).
•
Processor Intel Core i7 (4 cores 8 threads) 3 GHz or higher.
•
8 GB Ram.
•
Hard drive 1TB.
•
DVD recorder.
•
Graphics card with the following specifications (ex : GTX 9 Go or NVIDIA® QUADRO®
M2000):
-
chipset Nvidia
-
Global memory ≥ 4 GB
-
Capability (=architecture) ≥ Maxwell.
•
Operating System Windows 10 - 64bit.
6.3 Software
The equipment Graphical User Interface can be run with the software provided with the machine
or integrated in a third party imaging and database software that complies with the following
specifications: it has to be CE marked as medical device of class IIa and integrate the equipment
SDK according to what stated in the document PANOW3D API programmer's guide Vn (n is the
document revision), contact Owandy Radiology to have the latest revision of the programmer's
document.
The 3D exams can be viewed with any software that can import, view and manage 3D volumes
saved in DICOM slices with the following maximum dimensions:
•
Normal resolution full volume: 532 slices, 492x492 pixels per slice, 12 or 16 bits, for a
total of 484 kB/slice
•
Full resolution 80x50 volume: 542 slices, 984x984 pixels per slice, 12 or 16 bits, for a
total of 968 kB/slice.
6.4 I-MAX 3D
– PC communication
The communication between I-MAX 3D and computer is carried out with a LAN connection based
on a TCP/IP protocol and a point to point connection between the detector and computer using
an UDP protocol on a gigabit Ethernet.
In order to properly operate the unit, follow carefully the instructions reported in the Service
Manual at paragraph 7.6.
The system is provided with 2 Ethernet cat 5e cables in order to permit the PC connection. In
case of replacement, cables of the same or superior category have to be used.
If the communication between I-MAX and PC is not properly set problems in unit connection
causing impossibility of acquisition or loss of frames causing distortion and artefacts on the
images can occur.
Note
I-MAX 3D is not intended to transmit or receive information to/or from other equipment
through network/data couplings.