User Manual
– 3D exam
NIMXEN020I
Owandy Radiology SAS
84
13.
3D EXAMS
Note
Especially with pediatric patients evaluate the opportunity to select a smaller Field Of
View in order to reduce the dose delivered to the patient.
13.1 3D Full Dentition
The acquired volume is an 85x93 mm FOV centered to the
patient dentition.
In case the collimator kit code 6604061200 is present, the
acquired volume will change to 80x80 mm FOV.
13.2 3D SINUS
The acquired volume is an 85x90 mm FOV centered to the
maxillary sinus region.
In case the collimator kit code 6604061200 is present, the
acquired volume will change to 80x80 mm FOV.
13.3 3D TMJ
The acquired volume is an 85x90 mm FOV centered to the
temporo mandibular joint right or left, accordingly with the user
selection.
In case the collimator kit code 6604061200 is present, the
acquired volume will change to 80x80 mm FOV.
13.4 3D Single Jaw
The acquired volume is an 85x50 mm FOV centered either on the
maxillary or mandibular arch, depending on the user selection.
In case the collimator kit code 6604061200 is present, the
acquired volume will change to 80x50 mm FOV.
Two acquisition modes are available: Standard definition (175 µm
voxel size) for a standard resolution 3D acquisition, and High
definition (87.5 µm voxel size) for the enhancement of the finest
details.
Summary of Contents for i-max touch 3D
Page 1: ...EN USER MANUAL I MAX 3D NIMXEN020I June 2022...
Page 2: ......
Page 6: ......
Page 119: ...User Manual NIMXEN020G Owandy Radiology SAS 109...