User Manual
– Error message
NIMXEN020I
Owandy Radiology SAS
95
15.
ERROR MESSAGES
The error messages are divided into different areas that can be distinguished by the error number;
the following table contains the different errors with meanings.
Main MCU board
Code
Error description
000 / 001
Internal MCU error
400 ÷ 411
MCU
– DSPU communication fault (CAN Bus)
MCU EEPROM configuration
Code
Error description
100 / 101
Configuration area parameter doesn't match the expected one
102
Wrong version number in configuration area
103 / 104
Timeout error occurred during an EEPROM erase/write operation
Rotation Motor
Code
Error description
200
Zero position optical sensor of rotation axis always activated
201
Zero position optical sensor never activated
202 / 203
Zero position optical sensor of rotation still active after exiting from zero
sensor
204
Unexpected activation of rotation optical sensor
205
Timeout on rotation
Y translation motor
Code
Error description
240
Zero position micro Y always active
241
Zero position micro Y never active
243
Timeout on Y axes
Disk collimator
Code
Error description
260
Collimator timeout on zero positioning
Chin Rest
Code
Error description
265
Zero position optical sensor always activated
266
Zero position optical sensor never activated
268
Timeout on movement
Hardware key board (U.I.C.)
Code
Error description
270 / 271
Hardware key fault
X-Ray Controls
Code
Error description
360
RX button pressed on start-up or before exam
362
RX button released during emission