User Manual
–Panoramic image assessment
NIMXEN020I
Owandy Radiology SAS
102
17.2 Patient positioning errors
17.2.1
Turned head
Figure 37
Problem
The patient's head is turned to one side (left or right) in the
mid-sagittal plane.
Effects
Condyles are different in size.
The ramous on one side is much wider that the other one.
Asymmetric spine compensation.