Ottobock | 69
13.2.3 Error messages while establishing a connection with the cockpit app
Error message
Cause
Correction
Device not found
The connection could
not be established as no
component with the
entered serial number
could be found.
Compare the serial number entered with
that of the component, then try to establish
the connection again.
Component was con
nected to another
device. Establish con
nection?
The component was
connected to another
device/remote control
To disconnect the original connection, tap
the "
OK
" button.
If the original connection is not to be dis
connected, tap the "
Cancel
" button.
A current connection to
the prosthesis was inter
rupted
Check the following points:
•
Distance from the prosthesis to the
device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched
on? (see Page 60)
•
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
13.2.4 Status signals
Battery charger is connected
LED on
power
supply
LED on bat
tery charger
Event
Power supply and battery charger operational
Battery charger disconnected
Beep
signal
Vibration
signal
Event
1x short
1x short
Self-test completed successfully. Product is operational.
Battery charge level
Battery char
ger
Battery is being charged, battery charge level is less than 50%
Battery is being charged, battery charge level is over 50%
Battery is fully charged (or connection with product is interrupted).
Take note of the beep signals to differentiate between these states.
When the battery charger is connected or disconnected, a self-test is conduc
ted and confirmed by a single beep and vibration signal.
The battery is fully charged if this beep signal is heard.
If no beep signal is heard, the connection to the product is interrupted.