Ottobock | 67
Beep signal
Vibration
signal
Additional action per
formed
Event
2x short
2x short
Weight placed on prosthetic
leg and leg kept still for 1
second
Switching to MyMode 1 (mode 2)
carried out.
3x short
3x short
Weight placed on prosthetic
leg and leg kept still for 1
second
Switching to MyMode 2 (mode 3)
carried out.
13.2.2 Warnings/error signals
Error during use
Beep signal
Vibration signal
Event
3x long
Battery charge level under 25%
5x long
Battery charge level under 10%
10x long
10x long
Battery charge level under 5%
Beep signal
Vibration signal
Event
Required action
1 x short at interval of
approx. 5 seconds
Overheated hydraulic
unit
Reduce activity
10 x short
10 x long
Battery charge level
under 5 %
Charge battery
5 long beep signals
when malfunction
occurs
Error of moderate
severity
:
e.g. a sensor is not
operational
Switch to safety
mode.
Restricted walking pos
sible.
The product must be
inspected immediately by
a prosthetist.
5 x short at interval of
5 minutes
Active safety mode
Restricted walking pos
sible.
The product must be
inspected immediately by
a prosthetist.
30 x long
Intermittent for
5 minutes
Severe error
e.g. failure of valve
drives
Possibly no switching
into safety mode.
Restricted walking pos
sible.
Attempt to reset this error
by connecting/discon
necting the battery char
ger.
If the error persists, it is
no longer permissible to
use the product. The
product must be inspec
ted immediately by a
prosthetist.