60 | Ottobock
The following parameters can be modified:
Remote control
parameter
Adjustment
software
range
App/remote
control
adjustment
range
Meaning
Pitch
1000 Hz —
4000 Hz
1000 Hz —
4000 Hz
Pitch (frequency) of beep signal for
confirmation tones
Volume
0 — 4
0 — 4
Volume of beep signal for confirma
tion tones
Heel Resist.
10 — 60
± 20
Plantar flexion damping.
How fast the forefoot lowers itself dur
ing heel load.
Rollover Resist.
110 — 170
± 10
This parameter defines how smooth
rollover is.
Stair Function
ON — OFF
ON — OFF
Switching on this function increases
the rollover angle when walking down
stairs. This function needs to be
enabled in the adjustment software.
8.4.3 Overview of adjustment parameters in MyModes
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific
motion pattern such as cross-country skiing. Damping behaviour is not automatically controlled
and adjusted in MyModes.
The following parameters can be modified in MyModes:
Parameter
Adjustment
software
range
App/remote
control
adjustment
range
Meaning
Heel Resist.
0 — 195
± 20
Plantar flexion damping.
How fast the forefoot lowers itself dur
ing heel load.
Rollover Resist.
0 — 195
± 10
Dorsiflexion damping.
How easily the value of the "Stop
angle" parameter can be reached, in
other words how strong the resistance
is to reaching the value of the "Stop
angle" parameter.
Stop angle
-200 — 200
± 10
displayed in
0.1°
Ankle angle from which motion in the
rollover direction (in direction of dor
siflexion) is locked.
8.5 Turning Bluetooth on the prosthesis on/off
INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the cockpit app.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down or by
connecting/disconnecting the battery charger. Bluetooth is then turned on for approx. 2
minutes. During this time, the app must be started and used to establish a connection. If