Sym
bols on
the
control
panel
Beep
signal
Vibra
tion sig
nal
Event
Required action
—
—
The product switches itself off. This
occurs automatically in the following
cases:
•
The
button on the control pan
el was pressed and held for
longer than approx. 3 seconds
while product was switched on.
•
After the
symbol lights up.
•
After disconnecting the battery
charger, if the product was
already switched off before con
necting the battery charger.
•
Charge the battery.
•
If desired, switch on the product with
the
button.
—
—
Ongoing
Total failure
Electronic control no longer possible.
Safety mode active or undetermined
valve state. Undetermined product
behaviour.
1. Press and hold the
button on the
control panel until the vibration signal
stops (approx. 10 seconds), com
pletely switching off the product.
2. If the vibration signal recurs after
switching on, attempt to reset this error
by connecting/disconnecting the bat
tery charger.
3. If the vibration signal recurs, use of the
product is prohibited. The product
must be inspected by an authorised
Ottobock Service Center. For disas
sembly of the joint unit, see the section
“Assembling/disassembling the joint
unit on the orthosis” (see Page 79)
17.2.2 Error messages while establishing a connection with the cockpit app
Error message
Cause
Correction
Component was connec
ted to another device.
Establish connection?
The component was connec
ted to another device
To disconnect the original connection, tap the "
OK
"
button.
If the original connection is not to be disconnected,
tap the "
Cancel
" button.
Mode change failed
An attempt was made to
switch to a different MyMode
while the component was in
motion (e. g. while walking)
For safety reasons, switching MyModes is only per
mitted when components are at rest, e.g. while
standing or sitting.
A current connection to the
component was interrupted
Check the following points:
•
Distance from the component to the device
•
Charge level of the component's battery
•
Bluetooth of the component switched on? (see
Page 98)
•
Briefly press the button on the control panel
to make the component "visible" for 2 minutes.
•
Component switched on? (see Page 97)
•
If multiple components were stored, was the
correct component selected?
•
Is the component still connected to another
device, and is this connection still active?
110