INFORMATION
Bluetooth of the component must be switched on to change the orthosis settings. Briefly press the
button on
the control panel to check.
If the symbol is not lit up, Bluetooth has to be activated by a long press of the button on the control panel.
Information for changing the orthosis settings
•
Before changing settings, always check the main menu of the Cockpit app to make sure the correct component
has been selected. Otherwise parameters could be changed for the wrong component.
•
It is not possible to change orthosis settings nor to switch to a different MyMode while the orthosis battery is
being charged. Only the status of the orthosis can be called up. Instead of the
symbol, the
symbol
appears in the bottom row of the screen in the Cockpit app.
•
If the settings of a MyMode are to be modified, one must first switch to this MyMode.
10.4.1 Changing the orthosis settings using the Cockpit app
1) Once the component is connected and in the desired mode, tap the
symbol in the main menu.
→
The navigation menu opens.
2) Tap the "
Settings
" menu option.
→
A list appears with the parameters for the currently selected mode.
3) Change the setting of the desired parameter by tapping the "
<
", "
>
"
symbols.
INFORMATION: The setting configured using the adjustment app
is marked and, after the setting has been changed, can be
restored by tapping the "Standard" button.
10.4.2 Overview of adjustment parameters in basic mode
The parameters in basic mode describe the dynamic behaviour of the orthosis in the gait cycle. These parameters
act as basic settings for automatically adjusting the damping behaviour to the current motion situation (e.g. ramps,
slow walking speed, etc.).
The stance function and/or the sitting function can also be activated/deactivated. Further information on the stance
function (see Page 84). Further information on the sitting function (see Page 93).
The following parameters can be modified:
Parameter
Adjustment
range of the
O&P profession
al via the adjust
ment app
Cockpit app
adjustment
range
Meaning
Stance flexion resistance
120 to 190
+/- 10 of the con
figured value
Resistance against flexion motion e.g. when
walking down stairs or while sitting down
Stance function
0/Off - deactiv
ated
1/On - activated
Activation/deactivation of the stance function.
This function needs to be enabled in the
adjustment app.