8
Quality Tests
From the
Quality Tests
menu you can test the probe and cables connected to AccuScreen.
We recommend that you test the probe and cables at the beginning of each work day.
Note
•
You should always comply with local regulations and routines.
8.1
Performing Quality Tests
1. Press the
Quality Tests
button on the
Home
screen.
2. From the
Quality Tests
menu you can access the quality test types available in
AccuScreen:
•
Probe
•
Electrode cable
•
Ear coupler cable
•
Combined
(Electrode cable - Ear coupler cable)
Combined (Electrode Cable - Ear Coupler Cable)
8.2
Probe test
Every day before you start testing on patients, you should test the probe to make sure that it functions correctly.
Caution
•
Make sure that the probe tip has been cleaned and disinfected before you insert it in the test cavity. See
Cleaning and disinfecting the built-in test cavity
.
You can start the probe test in two ways:
•
, simply by inserting the probe in the test cavity of AccuScreen.
•
Otometrics - MADSEN AccuScreen
65
8 Quality Tests