8.2.2
From the Quality Tests menu
1. Insert the probe tip without eartip in the test cavity on the device, or in an
external test cavity.
2. Press the
Probe
button on the
Quality Tests
screen.
3. For more information, see
.
8.3
Electrode cable test
Every day before you start testing on patients, you should test the electrode cable to make sure that it functions correctly.
1. Clip the electrode clips onto the metal bars of the ABR Tester.
2. Press the
Electrode cable
button on the
Quality Tests
screen.
If the electrode cables function correctly, the message
Electrode cable OK
appears.