The
Pause
button changes into a
Start
button.
Pressing the
Pause
button will store data registered up to this point in time. This
will prevent registered data from being influenced by noise or movements.
By default, the test can be paused for max. 5 minutes each time you pause it. (The
maximum length of the pause can be configured in AccuLink.)
•
Press the
Start
button to resume the test.
Stopping a test
•
If needed, press the
Stop
button to stop the test. The test will be saved as an
incomplete test.
Test details
Shows which ear is being tested.
Single ear testing
Binaural testing
ABR
Shows the ABR response.
EEG
Is an indication of the amplitude of the total registered potential. The lower this
value is, the easier the test will be. If the electrode signals are significantly high (dis-
turbances, muscular artifacts) the data measured during this time will be ignored.
Impedance
These impedance values are updated and shown every 20 seconds.
52
Otometrics - MADSEN AccuScreen
5 Testing with AccuScreen