Modem:
IR data transmission to docking station - RS232 interface from docking station to
modem
Transport and storage environment
Temperature range:
-20 - +60°C (-4 - 140°F)
Humidity range:
10 - 90 % rel., non-condensing
Air pressure
500 hPa to 1040 hPa
Operating environment
Mode of operation:
Continuous
Temperature range:
10 - 40°C (50 - 104°F)
Humidity range:
30 - 90 % rel., non-condensing
Air pressure
500 hPa* to 1060 hPa
500 hPa to 1060 hPa
*At locations where the normal air pressure is below 800 hPa (at altitudes of
more than 2000 meters), it is recommended to recalibrate the OAE probes
and ABR ear coupler cable (if used).
Warm-up time
< 20 seconds.
Note
•
Should be extended if MADSEN AccuScreen has been stored in a cold environment.
Disposal
MADSEN AccuScreen can be disposed of as normal electronic waste, according to WEEE and local regulations.
Essential performance
MADSEN AccuScreen has no essential performance and accordingly, the applicable requirements are as stated in the fol-
lowing:
•
Otoacoustic emissions as defined by EN 60645-6, Type 2
•
Auditory evoked potentials as defined by EN 60645-7, Type 2
•
Basic safety as defined by IEC 60601-1
All information required by IEC 60601-1-2:2007, #5.2.2.1-#5.2.2.10 is available in the MADSEN AccuScreen User Manual.
Standards
Otoacoustic emissions:
EN 60645-6, Type 2
Auditory evoked potentials:
EN 60645-7, Type 2
Otometrics - MADSEN AccuScreen
97
App. 3 Technical Specifications