If needed, press the
Skip
button to skip a frequency.
The number of frequencies being tested is configurable. See the AccuLink User Manual.
•
If needed, press the
Stop
button to stop the test. The test will be saved as an
incomplete test.
Test details
Shows which ear is being tested.
Bar chart
Noise
Registered DPOAE response: Frequency passed
DPOAE frequency being tested, skipped or referred
The height of the bars represents the noise/response amplitude.
The test frequencies are shown below the bar chart.
The test protocols are configurable (see the AccuLink User Manual).
In a standard DPOAE test the 4 frequencies are
•
5 kHz
•
4 kHz
•
3 kHz
•
2 kHz
where 5 kHz is tested first.
The default
Pass/Clear Response
and
Refer/No Clear Response
criterion is 3/4. The test stops automatically if 3 fre-
quencies reach a
Pass/Clear Response
result or 2 frequencies receive a
Refer/No Clear Response
result.
DPOAE
Response at current test frequency.
Noise
Current noise level. Make sure the noise level is as low as pos-
sible.
Progress
Shows the time elapsed for the current test frequency. If a
Pass/Clear Response
for the current frequency is not achieved
the test frequency will time out. Max. time depends on envir-
onmental noise conditions. If test conditions are optimal but
there is no response measured, then the frequency will time out
before the progress bar reaches the end.
Otometrics - MADSEN AccuScreen
43
5 Testing with AccuScreen