8.5
Combined (Electrode Cable - Ear Coupler Cable)
Every day before you start testing on patients, you can test the electrode cable and ear coupler cable to make sure that
they function correctly. The combined test consists of two parts:
1. An automatic test to check whether there is a cable breakage on either the ear coupler cable or the electrode cable.
2. A manual listening test to make sure the ear coupler adapters are not clogged.
Automatic test
1. Clip the electrode clips onto the metal bars of the ABR Tester.
2. Insert the ear coupler adapters into the sockets at each end of the ABR
Tester.
3. Press the
Combined
button on the
Quality Tests
screen.
If the message
All tests OK
appears, the electrode cables and ear coupler
cables function correctly.
If the message
Test FAILED
appears, one or several electrode cables and/or
ear coupler cables do not function correctly.
4. Press the
OK
button to confirm the test result.
70
Otometrics - MADSEN AccuScreen
8 Quality Tests