8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
87
Figure 90. Detective-EX Reports Tab.
Figure 91. Detective-EX About Tab.
8.3.6. Reports
This tab (Fig. 90) displays the
reports available for the current live
spectrum or spectrum file, including
the ID, Found Nuclides, Suspect
Nuclides, Top Lines, and (for instru-
ments so equipped) GPS reports. You
must be running MAESTRO or other
ORTEC spectroscopy applications.
When using the Detective-EX as an
MCA controlled by the host computer,
be sure to click the Refresh button for
each report to ensure you are seeing
the most current data.
For more information on the contents
of these reports, see Section 5.1.6.
8.3.7. About
This tab (Fig. 91) displays hardware
and firmware information about the
Detective-EX as well as the data
Acquisition Start Time (Sample
description is not used). In addition,
the Access field shows whether the
MCB is currently locked with a pass-
word by MAESTRO. Read/Write
indicates that the MCB is unlocked
and Read Only means it is locked.
See the note concerning passwords on
page 82, at the end of Section 8.1.
Use the Firmware Revision field to
determine both the type of Detective-
EX (regular or -100 model, with or without neutron detector) and firmware version. This para-
meter is formatted
DETX-nnn
, where
X
is
C
for the regular Detective-EX,
D
for the Detective-
EX-100,
F
for the regular Detective-DX, and
G
for the Detective-DX-100; and
nnn
is the
firmware version. In the accompanying illustration,
DETF-304
indicates a regular Detective-DX