Figure 85. Detective-EX Amplifier Tab.
and Identification Mode messages for spectra downloaded from the Detective-EX, move from
tab to tab. Click on Close when finished.
Note that as you enter characters in the data-entry fields, the characters will be underlined until
you move to another field or until 5 seconds have lapsed since a character was last entered.
During the time the entry is underlined, no other program or computer on the network can
modify this value. The changes you make on most property tabs take place immediately. There is
no cancel or undo for the dialog.
NOTE
In MAESTRO, the Download Spectra and View ZDT Corrected commands on the
Acquire menu and the Smooth and Strip commands on the Calculate menu are
disabled for this instrument.
8.3.1. Amplifier
Figure 85 shows the Amplifier tab. This
tab displays the settings for the Coarse
and Fine Gain, which are factory preset
and cannot be changed. The fine gain
changes when the unit is recalibrated.