8. USING THE DETECTIVE-EX WITH ORTEC SPECTRUM ANALYSIS PROGRAMS
83
To prepare this calibration file, load a Detective-EX calibration spectrum into GammaVision or
ISOTOPIC. Use the software application’s Calibration Wizard to obtain the additional
efficiency calibration, keeping the existing energy calibration. After the efficiency calibration is
computed, save the complete set of calibrations as a
.CLB
file.
To add the efficiency information to an existing
.SPC
file (typically from a standalone
Detective-EX):
!
Transfer the
.SPC
file from the Detective-EX to the computer (see Chapter 6).
!
Recall the
.SPC
file in GammaVision or ISOTOPIC.
!
Load the
.CLB
file containing the proper efficiency information. This can be done either
with the application’s Recall Calibration command; or by specifying the
.CLB
file as the
calibration override in the GammaVision analysis options (
.SDF
) file or the ISOTOPIC
configuration.
!
Save the
.SPC
file with or without further analysis.
NOTE
If the
.CLB
file is specified in the GammaVision
.SDF
file or ISOTOPIC
configuration, the
.An1
file generated during analysis will contain the proper
calibration information but the original
.SPC
file (without the efficiency
information) will not change.
To load the efficiency information before acquiring a new spectrum (with the Detective-EX
attached to a computer):
!
Load the
.CLB
file either with the Recall Calibration command; or by specifying the
.CLB
file as the calibration override in the GammaVision
.SDF
file or ISOTOPIC
configuration.
!
Acquire a new spectrum and save it in
.SPC
format.
8.3. Detective-EX MCB Properties in MAESTRO
This section discusses the hardware setup dialogs you will see within MAESTRO when you use
the MCB Properties... command on the Acquire menu. The MCB Properties dialog contains all
of the instrument controls including acquisition presets, high voltage, hardware status indicators,
and amplifier settings. As noted in the introduction, except for the presets, the Detective-EX
acquisition parameters are factory-set and cannot be adjusted. To view the hardware parameters,