81
8. USING THE DETECTIVE-EX WITH ORTEC
SPECTRUM ANALYSIS PROGRAMS
The Detective-EX is intended for completely independent operation, with no computer required.
However, when you connect the Detective-EX CONTROL port to a computer running ORTEC
spectroscopy applications such MAESTRO or GammaVision, you can use it as a high-perform-
ance HPGe spectrometer/digital MCB.
Be sure to install the accompanying versions of C
ONNECTIONS
Driver Update Kit (P/N 797230)
and MAESTRO. These contain the most up-to-date drivers for the Detective-EX/DX-100T and
other ORTEC hardware. If you have any questions about software/hardware compatibility, con-
tact your ORTEC representative or our Global Service Center. The MAESTRO User’s Manual
contains complete instructions on software installation and configuration of the MCBs attached
to your computer. When installing C
ONNECTIONS
, choose the USB-Based Instruments selection
to install the correct driver for the Detective-EX/DX-100T.
8.1. Connecting the Detective-EX to a Computer
To enter MCA Mode, simply connect the Detective-EX to the computer via the rear-panel
CONTROL port. Within a few seconds the Detective-EX screen will change to the display
shown in Fig. 84 and all touchscreen features will be disabled. The unit can then be used as a
high-performance MCB in MAESTRO, etc.
The first time this Detective-EX is connected to a particular for use as an MCB, follow the
installation instructions in Chapter 6 for the host computer’s operating system.
IMPORTANT
When both the Detective and trans-SPEC applications are installed on your identifier,
the unit’s
operation depends on the application running when it is connected to the analysis computer.
If running the Detective application, the allowable changes to hardware settings within
MAESTRO are restricted. This includes the gain, rise time, calibration, and other settings that
will affect the location of gamma peaks (see Section for the allowable hardware settings). When
you disconnect the identifier’s CONTROL port from the computer, the unit automatically restores
all factory settings for the standalone Detective.
If the unit is running the trans-SPEC application, almost all hardware settings are adjustable (see
the trans-SPEC manual for complete details on the allowable hardware settings). Any changes
made in MAESTRO remain in effect when you disconnect from the computer and return to
standalone trans-SPEC operation.
Summary of Contents for Detective-DX-100T
Page 4: ...iv Intentionally blank ...
Page 10: ...x ...