Chapter 4 - Technical Specification
4.1.6 Table requirements
WARNING
Do not plug the scan head into a power block or powered table. The scan head must be
plugged directly into a wall socket.
The scan head should be positioned in the middle of the table width. The front of the scan
head should be inline with the front of the table to allow the patient to get into the correct
alignment position.
The chin rest should slightly overhang the edge of the table.
The scan head and touch screen must be positioned on a table that meets the necessary safety standards.
The table must support the weight of the scan head.
TABLE 13: Table size requirements
Dimension
Details
Width
887 mm (35 inches)
Depth
600 mm (24 inches) including chin rest
Height
725 mm (29 inches) to 1205 mm (47 inches)
Where wheelchair accessibility is required the table must meet wheelchair accessibility standards.
4.1.7 Battery
CAUTION
Do not remove the scan head covers. Only Optos trained technical personnel are
permitted to service the device.
The device contains a 3V Lithium coin battery.
4.1.8 Environmental considerations (ambient)
Do not operate the device in an environment that exceeds the environmental limits, see
TABLE 14: Environmental ranges
Status
Temperature
Relative Humidity
(Non-condensing)
Atmospheric Pressure
Normal use
+10°C to +35°C
75%
70kPa to 106kPa
Storage
-10°C to +50°C
95%
70kPa to 106kPa
Transport
-10°C to +50°C
95%
50kPa to 106kPa
Note
Device is DEHP free.
Page 62 of 72
Part Number: G110230/1ENG
English
Copyright 2019, Optos plc. All rights reserved.
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...