Chapter 2 - Introduction
No. Name of Part
Description
5
Status indicator
The color changes to show the status of the device, see
6
Touch screen
The touch screen monitor is used to interact with the system. It displays
alignment feedback and captured images.
7
Hand control
The hand control comprises buttons used to adjust the patient alignment and
capture images.
8
Image server
Runs the Optos
Advance
software and lets you manage the networked images.
The image server will also let you review and analyze patient images.
9
Customer network
Switches, Firewalls, Internet access, Review clients, Virtual image server (if
required)
10
Patient
environment
The image server, switch and any browser-based review clients or monitors must
be positioned outside the patient environment; more than 1.5 meters (5 ft) from
the scan head.
Note
The device also includes a user guide, power cable and ethernet cable not shown on this diagram.
A table may also be supplied.
2.6.2 Network specification and characteristics
The following network specification is recommended when connecting the image server to a Wide Area
Network (WAN).
TABLE 1: Network specification
Network specification details
Installation
Scan head and image server PC
Connection type
LAN TCP/IP ipV4
Connection speed
1 Gbps recommended
Connection security
All transport HTTPS by default
Connection ports (defaults)
443, 56620
Unused ports
Locked on scan head and image server PC
Part Number: G110230/1ENG
Page 23 of 72
Copyright 2019, Optos plc. All rights reserved.
English
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...