Chapter 3 - How to...
3.4.8.3 Cleaning
CAUTION
Wear powder-free gloves at all times when cleaning the device including the scan head
and main mirror. If the gloves become contaminated they must be replaced immediately.
In order to prevent any contamination always clean and disinfect parts of the equipment
which will be in contact with the patient, see
Adequate cleaning and disinfection is necessary to prevent disease transmission.
Thoroughly clean and disinfect the equipment surfaces that contact the patient and all
equipment surfaces likely to become soiled during use.
Improper cleaning methods or use of inappropriate cleaning and disinfecting agents can
result in the following:
l
Damaged equipment
l
Poor imaging performance
l
Increased risk of electric shock
Always follow the cleaning and disinfecting agents’ manufacturer’s instructions and
precautions; for storage, method of application, protective gloves, shelf life and disposal
to help assure effective and safe use of the product.
Do not use lint cloths, tissues, or other materials that may create dust, on or near the scan
head. Always allow cleaned areas to air-dry.
Do not use harsh detergent, abrasive cleaners or solvents that are not recommended by
the manufacturer.
Do not expose equipment parts to steam or sterilizers as this will damage the device.
Do not allow liquids to get into the equipment.
If you believe cleaning sprays or liquids have got into the equipment, disconnect the
power supply and have the equipment checked by Optos qualified service personnel.
Do not let the cleaning wipes come into contact with the inside of the instrument.
Do not use the equipment until you have confirmed it is safe to use.
3.4.8.3.1 General cleaning
CAUTION
Be careful not to damage the scan head covers when cleaning.
The equipment should be kept clean and dust free.
l
Logout, shut down and power off the device before cleaning the outer casing of the scan head.
l
Use a soft, slightly damp cloth to clean the plastic surfaces.
l
Ensure debris does not fall inside the device.
l
Use a glass-cleaning agent to clean the PC monitor and touch screen.
Page 52 of 72
Part Number: G110230/1ENG
English
Copyright 2019, Optos plc. All rights reserved.
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...