Chapter 4 - Technical Specification
Feature description
Compliance
Power Consumption
The touch screen must be powered through the
scan head and is included in these figures.
289-350VA
3.5A at 100-120Vac
1.75A at 200-240Vac
SMPS continuous
Touch screen voltage
+12V
4.1.2 Connections
These connections are not intended to be detached by the user.
FIGURE 16: Sample of the connection label
No. Description
1
Mains power cable and switch
l
Protective earth must be fitted.
l
U.K. Power Cable Fuse: 13A
2
Network cable
3
Hand control and Optos Service use
4
Touch screen
4.1.3 Scan head dimensions
The scan head, hand control and touch screen should be placed on a table which meets the table
specification, see
on page 62. There should be adequate clearance around the table to
operate the device.