
Chapter 1 - Warnings and cautions
1.5 EMC safety
CAUTION
In order to provide reasonable interference protection against radio communication
services, this device has been tested and complies with 47 CFR Part 15, Sub Part B of
the FCC (U.S.A.) rules and IEC/EN 60601-1-2 standard requirements at the premises of a
typical user, and the CISPR11 standard emission limits (CISPR11 Group 1 Class B),
Guidance and manufacturer declaration - electromagnetic emissions
Although compliant with all applicable international EMC standards, the equipment
generates, uses and can radiate radio frequency energy. The equipment may cause radio
frequency interference to other medical and non-medical devices and radio
communications nearby.
There is no guarantee that interference will not occur in a particular installation.
If this device is found to cause interference (which may be determined by turning the
equipment on and off), the user or qualified service personnel should attempt to correct the
problem by one or more of the following measures:
l
Relocate the affected device or devices
l
Increase the separation between the equipment and the affected device
l
Consult the Optos service representative for further suggestions
CAUTION
Unauthorized changes or modifications will invalidate the warranty of this device.
The manufacturer is not responsible for any interference caused by using:
l
Interconnect cables not supplied by Optos
l
Unauthorized changes or modifications to this equipment
Do not use devices which intentionally transmit RF signals (cellular phones, transceivers
or radio controlled devices) in the vicinity of this equipment, as it may cause performance
outside the published specifications, see
Recommended separation distances
Adhering to the distance separation recommended will reduce the disturbances recorded
at the image level, but may not eliminate all disturbances, see
on page 66. However, when installed and operated as specified herein the
system will maintain its essential performance by continuing to acquire, display and store
diagnostic quality images safely.
It is recommended that the image server is not used adjacent to or stacked with other
equipment.
The use of accessories and cables other than those specified may result in degraded
electromagnetic compatibility of the system.
1.6 Device failure
WARNING
In the unlikely event you hear a loud noise from the device, see smoke or smell burning,
stop imaging and isolate the device by unplugging.
If you suspect the device may be faulty see
If the fault still persists, please contact your Optos representative, see
Part Number: G110230/1ENG
Page 15 of 72
Copyright 2019, Optos plc. All rights reserved.
English
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...