Chapter 3 - How to...
A 3V Lithium coin battery is embedded in the hardware for the device and does not require to be changed by
users. If there is a problem, please contact your Optos representative, see
3.4.8.5 Cable management
WARNING
Do not plug the scan head into a power block or powered table. The scan head must be
plugged directly into a wall socket.
To prevent electric shock:
l
Only use cables supplied by Optos.
l
Cables should not be extended or altered.
l
Mains supply must be earthed.
l
The device may only be serviced by Optos trained personnel.
l
Care must be taken never to touch exposed parts of the device while in physical
contact with a patient.
If the mains cable is damaged do not use the system and contact your Optos
representative, see
CAUTION
You must regularly inspect the equipment. If there is any damage please contact your
Optos representative, see
Ensure the mains socket is easily accessible.
3.4.9 Troubleshooting
If you have any issues using your device please refer to the following table and instructions.
Additional information and resources are available on the Optos website. Go to
and register to
access the information and resources.
Detailed instructions on using Optos
Advance
can be found in the help files for each application.
Note
If you require further assistance, please contact your Optos representative, see
When contacting Optos you will need to supply your site number, see
What information do I need when
TABLE 8: List of problems and solutions
Problem
Possible cause
Solution
Capture mode is
disabled.
Failure of the system self-test
disables capture.
Open the diagnostics page and run the self-test
feature.
Cannot access the
user login list on the
scan head.
Password error, image server
is not switched on or scan
head is not configured with
the correct image server IP
address.
Check your user name and password, see
Check the image server is switched on and running,
see
How do I restart the image server?
Check the image server network address and ensure
the scan head is on the same network and configured
to connect to the correct address.
Part Number: G110230/1ENG
Page 55 of 72
Copyright 2019, Optos plc. All rights reserved.
English
Summary of Contents for P200T E Series
Page 1: ...optos com USER GUIDE Optos OCT Ophthalmoscope P200TxE ...
Page 4: ...Page 4 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 10: ...Page 10 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 70: ...Page 70 of 72 Part Number G110230 1ENG English Copyright 2019 Optos plc All rights reserved ...
Page 72: ...G110230 1ENG ...