LBL-2008-41-v5 OptoMonitor 3 IFU_en Rev 2021-08
96
|
P a g e
14
OPTOMONITOR 3 SYSTEM LOG – DISTAL INPUT VALUE
Dear valued customer,
The OptoMonitor 3 System log is used for recording the OptoMonitor 3 Distal Input value (diagnostic
value) on the initial installation. We recommend that you fill out the log and keep it in a file for
reference for your system maintenance. This log file serves as a comparative baseline for your future
maintenance purpose.
Parameter
Descriptions
Typical signal with
Reference
Probe
Poor signal
Lamp (%)
Lamp driving level
70%
Contrast (%)
Optical signal contrast visibility level
50%
< 35%
SNR
Signal to noise ratio
> 4000
< 2000
Gain
Signal Maximal Amplitude
1.7
> 2.20
Δ nm
Difference in nm between the Distal Input
value and the Maintenance reference probe.
>± 8 nm
Authorized staff name
Date of Verification
Optical Unit SN #
Display Unit SN #
Atmospheric pressure (external
source) (unit)
Lamp level (%)
Lamp Shunt (mV)
Vpeak (mV)
Vsignal (V)
Vtemp (mV)
Gain
Mano (mmHg)
Mano Supply (V)
Contrast (%)
SNR
Reference probe SN#
Reference Probe value (nm)
Distal Input value (nm)
Distal Input Delta (nm)