Reviewing Data
Use the “Play” button to load curves and review the data.
Archiving Data
7.10.4.1
Saving Pressure Data to a USB Drive
1-
Insert a USB 3.0 drive into one of the USB ports along the bottom of the device. (NOTE: You may
need to unscrew the protective panel along the bottom of the device in order to insert the
drive).
2-
Press the “USB Export” button in the playback screen.
3-
Data will be stored on the USB key in .bmp, .data and .dcm format.
4-
To save data without patient information, press the “Anonymous copy to USB” button. This will
upload an anonymous version of the data (.bmp and .data).
5-
More than one file can be exported at a time. To select or deselect all files locally stored, press
the “Select All” button.
7.10.4.2
Uploading Procedures Data via DICOM Export
If the OptoMonitor 3 system is properly linked to the facility’s RIS and PACS, you can export procedures
data via DICOM:
1-
From the MAIN screen, press the “Patient” button to access patient information.
2-
Press the “DICOM List” button and select the patient whose data will be used for the case (see
section 6.1.2). Alternatively, manually enter patient information, filling the required fields to
begin a case. Press the green arrow in the bottom-right hand portion of the screen to return to
the MAIN screen.
3-
Perform one or multiple FFR, dPR, or dPR Pullback recording(s).
4-
After the desired recordings have been produced, press the “Archive” button in the playback
screen.
5-
Select one or more recording you wish to export. You can use the “Select All” button to select all
recordings.
6-
Press the “DICOM Export” button. Pressing this causes the system to send a screenshot
capturing all relevant graphical and numerical data for the recordings along with procedure-
specific data such as the FFR or dPR value. The images are then sent via DICOM to the PACS and
is now available for viewing/storing in patient data in the PACS.
If a recording’s data were to be modified by the user (e.g. the cursor position, the vessel, drug,
or step selection, the zoom value etc.), the affected recording’s status of exportation will be
reset. The “DICOM Export” button will need to be used again for its data to be resent to the
PACS.
Terminate Operation
Once procedure is completed, unit can be left turned on. Optional screen lock will activate after
specified delay. It is possible to turn unit OFF by pressing the power button on the Optical Unit for
approximately 5 consecutive seconds. The Display Unit can be turned off using the power button located
under the unit.