LBL-2008-41-v5 OptoMonitor 3 IFU_en Rev 2021-08
77
|
P a g e
8.3.4
Calibrating Aortic Output to Cathlab System
Step-by-step Aortic Output calibration is detailed in section 20.
Reviewing Optical Sensor Performance
From the maintenance screen, press the “Optical Sensor Performance” button located in the top-right
side of the grid.
Displays readings and data relevant to the performance of the OptoWire currently being used. If no wire
is plugged into the OpM3, if the OptoWire is malfunctioning, or if the OptoWire is not correctly plugged
into the OpM3, then no data will appear on this screen (as seen in the image bellow)
Once a functioning OptoWire is connected to the OpM3 system, this screen will populate with live data
relating to the performance of the OptoWire (image below). This screen displays to OptoWire’s Serial
Number, probe value in nanometers, plus the actual value in nanometers read by the Optical Unit and
the difference between the two (Delta).