LBL-2008-41-v5 OptoMonitor 3 IFU_en Rev 2021-08
88
|
P a g e
10
GENERAL CARE AND CLEANING
Cleaning
Keep your monitor, modules, cables and accessories free of dust and dirt. Clean the exterior surface on a
regular basis in compliance with your institution’s infection control and biomedical engineering
department.
After cleaning, check the equipment carefully. Do not use if you see signs of deterioration or damage.
Optical Unit and Display Unit
The Optical Unit can be cleaned using standard disinfectants used in hospital environment such as a low-
level disinfectant, solution of mild detergent and water, Isopropyl alcohol (70%) and any standard class
cleaner (for Display Unit surface only). Cleaning products should be as close to neutral PH as possible.
•
TURN OFF POWER to the Optical Unit but do not disconnect any interface cables.
•
Wipe the exterior with a soft lint-free cloth, lightly moistened with the cleaning solution and
wipe dry.
•
Take extra care when cleaning the screen of the monitor because it is more sensitive to rough
cleaning methods than the housing.
•
Do not allow water or cleaning solution to enter the measurement connectors. Wipe around,
not over, connector sockets.
•
Turn power ON again.
WARNINGS:
•
Disconnect all powered equipment from the power line before cleaning or
disinfecting its surface.
•
Do not disconnect other interface cables.
•
If you spill liquid on the system or accessories, or they are accidentally immersed
in liquid, contact your service personnel, OpSens or OpSens authorised service
personnel. Do not operate the equipment before it has been tested and
approved for further use.
CAUTIONS:
•
Do not autoclave or steam clean the device.
•
Do not immerse any part of the equipment or any accessories in liquid.
•
Do not pour or spray any liquids directly on the device or permit fluid to seep into
connections or openings.
•
Do not use any abrasive material to scrub the instruments since this could cause
damage to the system surface.
•
Make sure that every part of the unit is dry before reconnecting to the power
supply.
•
Never use bleach. Do not use strong solvents such as acetone or
trichloroethylene.