LBL-2008-41-v5 OptoMonitor 3 IFU_en Rev 2021-08
86
|
P a g e
Pressure scale
0-250, 0-200, 0-150, 0-100, -50-
300, -100-200
0-200
Time scale
3, 6, 15, 30
3
dPR enabled
Enabled, Disabled
Disabled
Prospective Equalization
Enabled, Disabled
Disabled
Averaging period
1, 2, 3, 4, 5
3
Pullback time scale
30, 45, 60, 75
45
Hide Patient Info
Enabled, Disabled
Disabled
Pd Delay
0-504
0
MAINTENANCE MENU Input/Output Calibration
Setting
Range
Default
PA Source
Transducer, AUX
AUX
Aortic Transducer Input – Gain
0.5-2.0
1.0
Aortic Transducer Output –
Output
0, 50, 100, 200, 300, -30,
Normal, 0-100, 0-200, 0-300, -
30-300
Normal
Aortic Transducer Output – Gain 0.5-2.0
1.0
Aux Input – Offset
-10.0-10.0
0.0
Aux Input – Gain
0.5-2.0
1.0
Pd Out – Output
0, 50, 100, 200, 300, -30,
Normal, 0-100, 0-200, 0-300, -
30-300
Normal
Pd Out – Offset
-10.0-10.0
0.0
Pd Out– Gain
0.5-2.0
1.0
MAINTENANCE MENU Security
Setting
Range
Default
Use locking password
Enabled, Disabled
Enabled
Password
alphabetical characters,
numerical, and symbols
Ops22
Lock After
Never, 10 min, 30 min, 60 min
Never
Maintenance Access Code
numerical
0750