Preventive Maintenance
Timeframe
Maintenance
Weekly
Clean the backplate free from any adhesive build up; this may need to be
performed more frequently if adhesive build up is excessive.
Monthly
Clean the lens with an air hose (50 to 70 PSI) or (345 to 480 kPa).
Do not touch the lens; this could leave fingerprints on the lens or misadjust
the focus.
Clean off the light lens using alcohol and a lint free cloth if required.
Annually
If you have the Barcode Verification feature, replace the Calibration Card
every year. A new calibration card needs to be purchased through
Omron
Microscan.
Omron
Microscan does not give a schedule as to when you need
to calibrate; this is left up to the customer. However, we recommend that you
calibrate no less than once a month or if the environment has changed in
some way, such as lighting, movement of the system, etc.
WARNING:
Do not attempt to clean the Calibration Card as it will destroy the
image. Keep the Calibration Card out of the sun and in a clean environment.
Use extreme care to avoid fingerprints.