77
SAMPLING TIME:
3 samples per second. Re-zero of CJC
and auto calibrate every 5 seconds
TEMPERATURE
COEFFICIENT:
<150ppm/°C of max. linearized range
(typical)
REFERENCE
CONDITIONS:
22°C ±2°C, 115/230 V ±15% after 30
minutes settling time
SENSORS
INPUT TYPES:
Thermocouple: J, K, E, R, S, N, B, J DIN:
RTDs: 2-wire Pt-100, 3-wire Pt100
0 to 20 mV, 4 to 20 mV, user selectable
EXTERNAL
RESISTANCE (TC):
100 ohms maximum
COLD JUNCTION
COMPENSATION:
0.05°/°C (20:1)
BURNOUT
PROTECTION:
Fault display, upscale (downscale is key
selectable)
COMMONMODE
REJECTION
:
140 dB, 240 V, 50/60 Hz
NORMAL MODE
NOISE REJECTION:
60 dB, 50 Hz
Summary of Contents for C9000A SERIES
Page 6: ...This page is intentionally blank iv ...
Page 76: ...70 NOTES ...
Page 90: ...84 NOTES ...