69
Set (-12°) in Function .35°
5. A span error entered in Function .35 immediately changes the
reading. Allow time to stabilize at T2. If an error exists correct
with Function .9. Then check at T1. If an error exists check
readings and calculations, repeat if necessary.
E2-E1
T2-T1
Formula: Function .35 =
x span (as Function .24)
(-5°)-(+2°)
200°-60°
Example: Function .35 =
x 250° (Function .24 span)
-7
140
Function .35
x 250 = -12°
Summary of Contents for C9000A SERIES