68
NOTE
Error polarity applies to CN9000A correction
EXAMPLE READS
CN9000A
404°
Ext. Meter
400°
Error
+4
Set (-4) correction at Function .9.
6.2.2 Sensor Span Adjust (Function .35)
Figure 6-2. Sensor Span Graph
Steps for providing correction where two temperatures require
differing amounts of adjustment.
1. Choose a temperature near the bottom of the normal operating
range and one near the top.
2. Run at the lower temperature T1 &, note the error E1 between
CN9000A and "Reference" reading.
3. Repeat at upper temperature T2 and note error E2.
Example
T1 reads
T2 reads
CN9000A
60°
200°
"Reference" 58°
205°
Error E1=+2
E2=-5°
4. Calculate, span adjustment for Function .35
Temperature That CN9000A Reads
Temperature
That
‘Reference’
Reads
Negative
adjustment
Positive
adjustment
Ideal (i.e.
readings
match)
Summary of Contents for C9000A SERIES
Page 6: ...This page is intentionally blank iv ...
Page 76: ...70 NOTES ...
Page 90: ...84 NOTES ...