6 Stereo measurement procedures (RX (stereo measurement function) only)
7
Write down the measurement results.
8
When confirming the measurement values after the stereo optical adapter is reg-
istered or selected, make sure that the measurement results are within ±5%.
When confirming the measurement values before and after measurement, com-
pare the measurement values before and after measurement and make sure
that both measurement values are almost identical.
If the measurement results differ significantly from each other, check that the optical adapter is
not loose and the lens is not dirty and, if required, re-register the stereo optical adapter. (see "6.3
Registering the new stereo optical adapter or re-registering the stereo optical adapter" (P. 54)).
●
NOTE
For forward-view optical adapter
For side-view optical adapter
2 mm
0.1 inch