62
6 Stereo measurement procedures (RX (stereo measurement function) only)
Measure the object distance at the position of
the cursor in the left image.
The object distance indicator shows the distance between the distal end of the optical adapter and
the measurement area by nine levels.
Up to 15 mm:
(Recommended measurement distance)
Up to 30 mm:
(Possible measurement distance)
30 mm and higher:
↑
Green
↑
Yellow
↑
Red
To move the cursor, tilt the [MEAS/ENTER] joystick in the desired direction.
Moving the cursor
Spot ranging screen
Finishing the spot ranging
Press the [LIVE/GAIN] button, or press the [ZOOM] lever toward [T] or [W].
Short-press the [MEAS/ENTER] joystick. The measurement screen is displayed.
Starting the measurement
NOTE
●
While checking the distance to the measurement area by using the spot ranging function, approach to the
measurement area up to the recommended measurement distance. Generally the closer measurement
object distance is, the more accurate measruement becomes.