IPLEX GAir
Message
Cause and recommended action
the external battery became too hot.
→Stop inspection temporarily, turn OFF the power of
the system to allow this instrument to cool down, and
turn ON the power again.
<OVER CURRENT (COMPRESSOR
MOTOR). PLEASE TURN OFF THE
POWER.>
The self-check function was activated to prompt
termination of inspection due to the abnormal motor of
the compressor unit in the drum unit.
→Stop inspection temporarily, turn OFF the power of
the system to allow this instrument to cool down, and
turn ON the power again.
<ABNORMAL PRESSURE
(COMPRESSOR UNIT). PLEASE TURN
OFF THE POWER.>
The self-check function was activated to prompt
termination of inspection due to the abnormal pressure
of the compressor unit in the drum unit.
→Stop inspection temporarily, turn OFF the power of
the system to allow this instrument to cool down, and
turn ON the power again.
<HIGH TEMPERATURE (COMPRESSOR
UNIT). PLEASE TURN OFF THE
POWER.>
The self-check function was activated to prompt
termination of inspection because the inside of the
compressor unit in the drum unit became too hot.
→Stop inspection temporarily, turn OFF the power of
the system to allow this instrument to cool down, and
turn ON the power again.
<IMAGER ERROR. PLEASE TURN OFF
THE POWER.>
The self-check function was activated to prompt
termination of inspection due to abnormal image
acquisition function.
→Stop inspection temporarily, turn OFF the power of
the system, and turn ON the power again.
<UNRECOGNIZED SCOPE UNIT IS SET.
PLEASE TURN OFF THE POWER.>
The power is turned ON when the scope unit other than
the scope unit for this instrument is connected to the
drum unit.
→Connect the scope unit designated for this
instrument.
8.1.2 Common problems
Problem
Cause and recommended action
Illumination does not light.
The [LIGHT] button is not turned ON.
→Turn ON the button.
The optical adapter is not attached.
127