TIP:
l
Each petal of the plate has a small protrusion
(K)
as shown in the picture. When the
protrusion(K) on the petal is gone due to the wear of the plate, it is time to replace the plate.
Replace the plate according to the wear degree.
l
If the O-ring has a crack or dropped out, replace the O-ring with a spare one.
l
When replacing the O-ring, remove the O-ring from the cap using the small screwdriver, etc.
3.12 Pre-operation / Post-operation inspections
NOTE:
l
When inspecting the insertion tube, be sure to hold it at the position behind the angulation
section. Otherwise, the angulation section may be damaged.
l
Also, be sure to check the inspection items after using the instrument.
Inspection location
Item
If abnormality is discovered
Overall
□
Are foreign objects such as dirt or sand
attached to the LCD monitor, insertion
tube or doors?
"Storage and maintenance"
(page 131)
□
Are there cracks or scratches on the LCD
monitor, insertion tube or doors?
Replacement required.
Contact Olympus.
□
Any irregularities such as damage to or
deformation of the exterior materials,
buttons, joysticks, or levers?
□
Any cuts, buckling or other abnormalities
on the cable?
Insertion tube,
angulation
section, distal end
□
Any looseness in materials other than the
angulation section?
"Storage and maintenance"
(page 131)
□
Any dirt or water droplets on the objective
lens of the distal end?
Continued use in this condition creates
the risk of damage to the optical adapter
and/or distal end.
□
Any build up of dirt on the optical adapter
68
Summary of Contents for IPLEX GAir IV9000GA
Page 2: ......