68
Chapter 5 Operation
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
All cutting modes (except BlendCut) are
supported by the High Power Cut Support
(HPCS). The HPCS optimizes the start of the
cutting procedure by applying high power to the
tissue to support immediate spark ignition and
reduce the risk of mechanical cutting.
1.
Checking the output settings
Confirm that the mode, the output power level and the effect displayed on
the touch-screen correspond with the procedure to be performed.
If no change is required for the output settings, go to chapter 5.5,
“Output of energy”.
If the output settings have to be changed, go to step no. 2 below.
2.
Choose the power and effect settings
If the touch-screen shows the “All screen”, press on the appropriate button
connected to the output socket which settings shall be changed e.g.
“MONOPOLAR 1 button” (see figure 5.2).
The display changes from the “All screen” to the “Set screen”.
When the “Set screen” is displayed, the output power level and effect
settings can be changed.
Set the output power level by pressing the “Plus button” / “Minus button”
and the effect setting by pressing the “Toggle button”. The impact of
power level and effect setting on tissue for all modes is described in
table 5.1, “Overview of all modes”.
Figure 5.2
All screen
Set screen