Chapter 8 Troubleshooting
113
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
8.2 Error screen, codes and measures
Follow the troubleshooting advices in this chapter, to identify or correct failures.
The error window is configured as shown in figure 8.1.
Figure 8.1
If an error occurs (see figure 8.1):
An error window will appear and an alarm signal is audible.
A short message with the error code, error title and a description of the
remedial action will be displayed.
The error code consists of an error number shown under the “caution”
symbol.
Depending on the error priority, the condition of the audible signal and the
“caution” symbol are different (see table 8.1).
Proceed with the described remedial action.
The error window disappears after a few seconds, if the error is cleared.
If the error window is still displayed, the error is not cleared. Proceed with
the next remedial action if available.
Table 8.1: Error priorities and the corresponding indicator symbol condition
Error category
Error condition priority
Indicator (“caution”)
symbol condition
High priority
Immediate user response
is required
Flashes in red color
Medium priority
Prompt user response
is required
Flashes in yellow color
Low priority
Awareness of the user
is required
Constant on in yellow
color
Error code
Error title
Remedial actions
Error window
(Example: E002 Short circuit)
Caution symbol
Summary of Contents for ESG-400