Important Information – Please Read Before Use
9
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
electronic equipment. If an auxiliary computer
system
is
in
use
together
with
the
electrosurgical
generator
and
endoscopic
imaging techniques, the image on the monitor
might freeze or blackout. Follow the instructions
in
“Electromagnetic
Compatibility
(EMC)
information”
in
the
Appendix
regarding
electromagnetic ambient conditions.
Never loop the cords (active cord, bipolar cord,
neutral electrode cord) or bundle cords together
with cords belonging
to other medical
equipment. The high frequency signals or spark
discharge
noise
generated
by
the
electrosurgical generator may interfere with the
operation of other medical equipment.
Do not use the electrosurgical generator in a
location exposed to strong electromagnetic
radiation (microwave or short-wave medical
treatment equipment, magnetic resonance
imaging, radio or mobile phone equipment).
Electrosurgical generator malfunction can occur.
CAUTION
Unsuitable temperature and humidity
The electrosurgical generator should only be used
under the conditions as described in “Transport,
storage and operating environment” in the
Appendix. Use under other conditions may impede
normal performance and / or result in equipment
damage.
Accessories
WARNING
Mechanical stress
Do not apply excessive bending, straining, or
squeezing force to any cords. It may cause
malfunction.
CAUTION
Damaged and non-compatible accessories
The electrosurgical generator shall only be
used with compatible accessories. When
connecting accessories (cords, electrodes, HF
instruments) avoid output settings where the
maximum output voltage of the electrosurgical
generator may exceed the rated accessory
voltage (refer to “Mode characteristics”, “Output
characteristics” in the Appendix, and the
instruction manual of the accessory). For a list
of compatible neutral electrodes, refer to
“Specifications” in the Appendix. For compatible
Summary of Contents for ESG-400