Chapter 4 Inspection
61
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
4.6 Inspection of alarm system
WARNING
Be sure to inspect the alarm function before use. If
the alarm function is not normal, the electrosurgical
generator might fail to detect an equipment error
and this may result in unexpected burns,
perforation and / or bleeding.
Inspection of neutral electrode warning
1.
Make sure that no neutral electrode is connected to the neutral electrode
socket of the electrosurgical generator.
Try to activate any monopolar mode by pressing the corresponding
footswitch pedal. The contact quality monitor indicator for split neutral
electrodes illuminates red and the output cannot be activated.
2.
Confirm that an error window (E202, see figure 4.9) is displayed and an
audible signal can be heard.
3.
The audible signal stops and after a few seconds the error window (E202)
disappears.
Figure 4.9
4.7 Procedure after inspection
Switch off the electrosurgical generator with the power switch.
The illumination of the power switch is off.
The touch-screen is off.
The illumination of the push buttons is off.
The contact quality monitor indicator for split neutral electrodes is off.
Error window
(E202: Insufficient neutral
electrode contact)