Chapter 3 Installation and Connections
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
3.6 Connection of footswitch
There are two kinds of footswitches, the single pedal footswitch (optional) and
the double pedal footswitch. This section describes how to connect these
footswitches.
WARNING
Connect the footswitch plug securely to the
electrosurgical generator. Otherwise, output
may not be activated. In this case, the HF
instrument could cut the tissue mechanically,
which could cause bleeding and / or perforation
of the tissue.
The footswitch plug is not waterproof, and liquid
such as water must not get into the plug.
When
connecting
or
disconnecting
the
footswitch plug, always hold the plug. Pulling
the cable may result in damaging of the wires.
Do not connect products other than the
footswitch for this electrosurgical generator to
the footswitch socket. Otherwise, the footswitch
might not function and may cause patient injury
and / or damage of the equipment.