152
Appendix
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
Electromagnetic compatibility (EMC) information
The electrosurgical generator is intended for use in the environment specified
below. The user of the electrosurgical generator should assure that it is used in
such an environment.
Guidance and manufacturer’s declaration –
electromagnetic emissions
Emission test
Compliance
Electromagnetic environment – guidance
High frequency emissions
CISPR 11
Group 1
During standby the electrosurgical generator uses
high frequency energy only for its internal function.
Therefore, its high frequency emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
High frequency emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker
emissions IEC 61000-3-3
complies
The electrosurgical generator is suitable for use in all
establishments other than domestic, and may be
used in domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning:
This electrosurgical generator is
intended for use by healthcare professionals only.
This electrosurgical generator may cause radio
interference or may disrupt the operation of
nearby equipment. It may be necessary to take
mitigation measures, such as re-orienting or
relocating
the
electrosurgical
generator
or
shielding the location.
Length of power cord:
≤
4.5 m
CAUTION
The electrosurgical generator should not be
used adjacent to or stacked with other electrical
equipment, except devices which are intended
for this purpose and tested by the manufacturer.
The use of accessories which are not approved
by the manufacturer may result in an increase
of
electromagnetic
emissions
and
the
compliance with the stipulated limit values are
not guaranteed anymore.