156
Appendix
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
Alarm information
The electrosurgical generator complies with the IEC 60601-1-8: 2006. The
following table shows the priority relationship between errors and the output
tone frequency.
Error priorities, frequencies, duty cycles and pressure
levels of error tones
Error priority
Error codes
Duty cycle / frequency /
volume
High
E004, E007…E011, E018, E024…E031, E033…E037,
E039…E043, E048, E052, E054, E056…E061,
E063…E067, E074, E075, E081…E102, E116…E119,
E122…E130, E132, E133, E137, E142…E145, E150,
E165, E168…E178, E197, E198, E203…E211, E213,
E216, E217, E220…E387, E395, E400…E408, E410,
E412, E415, E419…E437, E439, E440, E443…E456,
E458…484, E490, E505…E514, E518…E538
Intermitting signal
(number of burst = 10) /
200 / 300 Hz /
≥
65 dB (A)
Medium
E012, E016, E017, E019…E023, E032, E038,
E044…E047, E049, E051, E053, E055, E062,
E068…E073, E078…E080, E103…E114, E120, E121,
E131, E134…E136, E138…E141, E147…E149,
E151…E154, E159, E166, E167, E179…E186,
E188…E196, E199…E201, E212, E215, E218,
E390…E394, E396…E399, E413, E414, E438, E457,
E488…E490; E515; E516
Intermitting signal
(number of burst = 3) /
200 / 300 Hz /
≥
65 dB (A)
Low
E001…E003, E006, E013…E015, E115, E187, E202,
E214, E409, E411, E416…E418, E441, E486
Intermitting signal
(number of burst = 2) /
200 / 300 Hz /
≥
65 dB (A)