Chapter 3 Installation and Connections
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
When using the electrosurgical generator with
the compatible ultrasonic generator and the
compatible high flow insufflation unit (UHI-2 or
UHI-3) simultaneously, contact Olympus.
1.
Preparation
Prepare the adapter for UHI-2/3 and the communication cable
The compatible high flow insufflation unit cannot be connected directly to
the electrosurgical generator. Use the adapter for UHI-2/3 (MAJ-1873) and
the communication cable (MAJ-1871 or 1872) to connect the UHI-2/3 with
the electrosurgical generator.
2.
Connecting the adapter for UHI-2/3
Connect the system plug of the adapter for UHI-2/3 to the system
connector on the rear panel of the compatible high flow insufflation unit
(see figure 3.2).
3.
Connecting the communication cable to the adapter for UHI-2/3
Connect the LINK-IN or LINK-OUT plug of the communication cable 10 m
to the LINK-IN or LINK-OUT socket of the adapter for UHI-2/3. After
connection, secure each plug by pushing it in while turning the screws on
both sides of the plug (see figure 3.2).
4.
Connecting the communication cable to the electrosurgical generator
Connect the other plug of the communication cable 10 m that has been
connected to the adapter for UHI-2/3 to the LINK-IN or LINK-OUT
connector of the electrosurgical generator. After connection, secure each
plug by pushing it in while turning the screws on both sides of the plug (see
figure 3.2).