Chapter 4 Inspection
ELECTROSURGICAL GENERATOR ESG-400
W7.091.601
4.2 Inspection of power
WARNING
If the electrosurgical generator fails to start up
(no indicators illuminates and the touch-screen
is off), confirm that the power cord is connected
securely to a grounded wall outlet and the AC
power socket
on the electrosurgical generator
and confirm that the grounded wall outlet is
powered. If it still fails to start up, remove the
power cord from the grounded wall outlet and
contact Olympus. Equipment damage or
malfunction may have occurred and fire or
electric shock can result.
If the output is activated and the output tone
can be heard but no output indicators
illuminates or the touch-screen
are off, stop the
procedure
immediately,
switch
off
the
electrosurgical generator. Otherwise, it may
cause perforation, bleeding and user and / or
patient burns.
Inspection of the power switch, touch-screen and push buttons
Switch on the electrosurgical generator with the power switch.
The power switch illuminates.
The touch-screen lights up.
The push buttons illuminates.
The contact quality monitor indicator for split neutral electrodes illuminates.