Chapter 7 Care, Storage and Disposal
109
W7.091.601
ELECTROSURGICAL GENERATOR ESG-400
2.
Disconnecting the accessories
Disconnect the accessories, footswitches and cables from the
electrosurgical generator.
3.
Disconnecting the power cord
Disconnect the power cord from the grounded wall outlet.
4.
Cleaning
To remove dust, dirt and non-patient debris, wipe the electrosurgical
generator (including the touch-screen) and footswitch using a soft, lint-free
cloth moistened with 70 % ethyl or isopropyl alcohol.
If the equipment and / or accessories are contaminated with blood or other
potentially infectious materials, first wipe off all gross debris using neutral
detergent, and then wipe its surface with a lint-free cloth moistened with a
surface disinfectant.
5.
Drying
After wiping with disinfecting alcohol, be sure to dry the electrosurgical
generator completely prior to use.
7.2 Storage
CAUTION
Do not store the electrosurgical generator in a
location exposed to direct sunlight, x-rays,
radioactivity, liquids or strong electromagnetic
radiation
(e.g.
near
microwave
medical
treatment
equipment,
short-wave
medical
treatment equipment, magnetic resonance
imaging equipment, radio or mobile phones).
Damage to the electrosurgical generator may
result.
Do not subject the electrosurgical generator to
strong impacts during storage. Doing so will
damage the electrosurgical generator.
1.
Storage location
Place the electrosurgical generator on a stable, level surface.
2.
Storage environment
Store the electrosurgical generator properly according to the environmental
conditions described in the “Transportation, storage and operating
environment” in the Appendix.