44
4-4 Advanced settings
1
Shading correction
You can correct the unevenness of brightness caused by the uneven
illumination to the even brightness.
To use the shading correction function, the magnification
of the objective of the microscope and the magnification
of the intermediate magnification changer must be
specified in advance. For details, see,"Setting the
microscope (p.84)".
The shading correction image is acquired per each combination of
the objective magnification and the magnification of the intermediate
magnification changer you specified. It takes approximately 1 minute
to acquire one combination.
Before acquiring the corrected image, place the specimen to bring it
into focus and then, remove the specimen in advance.
·
When acquiring the shading correction image, if there is
dust or dirt in the acquisition area, the image may not be
acquired properly. Make sure to confirm that there is no
dust or dirt in the acquisition area.
·
Up to 28 combinations of corrected images can be acquired.
If you attempt to acquire more than 28 combinations of
corrected images, an error message appears. In this case,
set the microscope again. For details, see,"Setting the
microscope (p.84)".
Select the [Acquisition] tab A .
NOTE
Acquiring the corrected image
NOTE
1
C
B
A
If the setting items of [Microscope] are not displayed, press B to
open them.
Press the [Device Settings] button C . The [Device Settings] dialog
box is displayed.
2
3